Comparing planning time, delivery time and plan quality for IMRT, RapidArc and tomotherapy
Abstract
The purpose of this study is to examine plan quality, treatment planning time, and estimated treatment delivery time for 5‐ and 9‐field sliding window IMRT, single and dual arc RapidArc, and tomotherapy. For four phantoms, 5‐ and 9‐field IMRT, single and dual arc RapidArc and tomotherapy plans were created. Plans were evaluated based on the ability to meet dose‐volume constraints, dose homogeneity index, radiation conformity index, planning time, estimated delivery time, integral dose, and volume receiving more than 2 and 5 Gy. For all of the phantoms, tomotherapy was able to meet the most optimization criteria during planning (50% for P1, 67% for P2, 0% for P3, and 50% for P4). RapidArc met less of the optimization criteria (25% for P1, 17% for P2, 0% for P3, and 0% for P4), while IMRT was never able to meet any of the constraints. In addition, tomotherapy plans were able to produce the most homogeneous dose. Tomotherapy plans had longer planning time, longer estimated treatment times, lower conformity index, and higher integral dose. Tomotherapy plans can produce plans of higher quality and have the capability to conform dose distributions better than IMRT or RapidArc in the axial plane, but exhibit increased dose superior and inferior to the target volume. RapidArc, however, is capable of producing better plans than IMRT for the test cases examined in this study.
PACS number: 87.55.x, 87.55.D, 87.55.de, 87.55.dk
I. INTRODUCTION
Intensity‐modulated radiation therapy (IMRT), intensity‐modulated arc therapy (IMAT) and tomotherapy (Tomo) are all advanced external beam radiation therapy treatment techniques that have been implemented for routine clinical use at different time points over the last 10 years.1 IMRT using a conventional linear accelerator equipped with a multi‐leaf collimator (MLC) was adapted for clinical use to treat prostate cancer in 1995 (despite the fact that IMRT using compensators was performed earlier).2 This was followed by the implementation of the technique for other treatment anatomical locations.(3,4) A recent meta‐analysis examined data from 56 clinical trials has suggested that IMRT can reduce toxicities as compared to non‐IMRT treatments; however, the data regarding local control and overall survival are inconclusive.5 With emerging data demonstrating the advantages of IMRT, future advancements in advanced radiation therapy delivery will include improvements in quality, efficiency, accuracy with image guidance, and ability to paint dose distributions.1 This paper investigates both improvements in quality and delivery efficiency of radiation therapy.
IMAT implemented with the gantry of the linear accelerator rotating during delivery along with MLC variations, was first proposed by Yu.6 Clinical implementation of arc therapy techniques was sparse, but treatments were performed for central nervous system, prostate, head and neck, whole abdominopelvic treatments, rectal cancer and endometrial cancers.(7–9) A major advance in IMAT was realized when Otto implemented his volumetric modulated arc therapy (VMAT) algorithm.10 VMAT uses a progressive sampling algorithm which starts with coarse gantry samples and then, throughout the optimization, the arc resolution is gradually improved. Without this algorithm, neighboring segments are highly restricted by the allowed leaf motion. VMAT obviates this restriction by allowing large leaf movements early in the optimization and more restricted leaf motion in the later stages.10 The optimization time is also greatly reduced. Otto's algorithm has been implemented by Varian (Varian Medical Systems, Palo Alto, CA, USA) and is marketed as RapidArc. In this implementation, the progressive sampling is achieved through five discrete “multi‐resolution” (MR) levels in which the number of segments increases from 10 to 177. Elekta (Elekta AB, Stockholm, Sweden) also have a product named VMAT which does not implement Otto's algorithm but uses a proprietary algorithm.11 Both Varian's and Elekta's implementations of arc therapy allow for dose rate variations.
Helical tomotherapy was first proposed by Mackie and is now commercially available from TomoTherapy (TomoTherapy Inc, Madison, WI, USA).12 Tomotherapy is a technique whereby a fan beam of radiation rotates around the patient who is translated through the bore of the tomotherapy machine as in conventional computed tomography. The beam trajectory follows a helical path during delivery and is modulated by a binary MLC. Treatments are optimized from 51 projections and can be conceptualized as IMRT beams delivered from 51 equally spaced angles.13 The first patient treated with helical tomotherapy was in 2002 and, since then, units have been sold worldwide.14
There is a need to characterize the differences in plan quality, planning time, and delivery time for IMRT, modern arc therapy, and tomotherapy. Previous studies have compared IMRT with tomotherapy, IMRT with arc therapy, and tomotherapy with arc therapy.(8,15–21) There is a single study that examines IMRT, arc therapy, and tomotherapy planning for five patients with benign intracranial lesions.22 The conclusions of this study are that all techniques are practically equivalent. The study recommends providing more challenging cases to evaluate the treatment techniques, and provides the primary motivation for this study. Furthermore, a recent publication by Bortfeld and Webb provided some theoretical considerations when considering the quality of dose distributions that can be achieved for IMRT, single arc IMRT, and tomotherapy based on a 2D phantom with an analytically derived solution.23 They conclude that a single arc that is delivered in less than 2 minutes may unduly compromise the plan quality for very complex cases, and feel that the plan quality for IMRT and single arc‐IMRT should be similar.
The goal of this study is to create IMRT, RapidArc and tomotherapy treatment plans for four virtual phantoms including the Brahme phantom23 and compare the treatment plan quality, optimization time and delivery time.
II. MATERIALS AND METHODS
A. overview of methods
A summary of the methodology for this paper has been included in Fig. 1 which refers to the different section headings, the major planning systems, and dose evaluation software that is used in this study. Data transferred between systems is listed either as an input or output arrow in Fig. 1.

B. Definitions of phantoms, contours and plan objectives
All virtual phantoms and contour sets were created within the Eclipse treatment planning system (Eclipse 8.6 pre‐release version with RapidArc planning capability, Varian Medical Systems, Palo Alto, CA). All phantoms were water equivalent and had a 10 cm radius and a length of 25.25 cm. An axial view of all of the phantoms including contours is provided in Fig. 2. Each phantom included an external contour, a planning target volume (PTV), and a number of organs at risk (OARs). The contours were extended in the third dimension by varying amounts. The superior‐inferior lengths of the PTVs and OARs are as follows: PTV length is 8.5 cm, OAR1 length is 6 cm for Phantom 1; PTV, OAR1, OAR2, OAR3 length is 5.25 cm for Phantoms 2 and 3; PTV length is 8 cm, OAR length is 7.5 cm for Phantom 4. The locations of the PTV and OARs were chosen because they represent geometries that are seen in the clinic or are well‐established test cases. Phantom 1 is similar to the American College of Radiology test case.24 Phantom 2 represents a prostate geometry, and Phantom 3 represents a head and neck geometry. Finally, Phantom 4 was the test case that was used in the analysis by Bortfeld.23
Planning objectives for all phantoms are listed in Table 1. The prescription dose for these treatment plans was 60 Gy in 30 fractions. The dose‐volume objectives were chosen to represent a significant challenge to each of the three techniques. The choice of importance values varied depending on the planning system that was used.

| PTV | OAR1 | OAR2 | OAR3 | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Min Dose (Gy) | Max Dose (Gy) | DVH Vol (%) | DVH Dose (Gy) | Max Dose (Gy) | DVH Vol (%) | DVH Dose (Gy) | Max Dose (Gy) | DVH Vol (%) | DVH Dose (Gy) | Max Dose (Gy) | |
| Phantom 1 | 60.0 | 65 | 50.0 | 13.3 | 20.0 | ||||||
| Phantom 2 | 60.0 | 65.0 | 50.0 | 10.0 | 20.0 | 50.0 | 10.0 | 20.0 | |||
| Phantom 3 | 60.0 | 65.0 | 50.0 | 10.0 | 20.0 | 50.0 | 10.0 | 20.0 | 50.0 | 15.0 | 30.0 |
| Phantom 4 | 60.0 | 65.0 | 50.0 | 15.0 | 30.0 | ||||||
C. Sliding window IMRT planning
Sliding window IMRT treatment plans were created within the Eclipse planning environment. Two plans were created consisting of either 5 (IMRT5) or 9 (IMRT9) beams using the 120 leaf Millennium MLC which has a 5 mm leaf width at isocenter. The choice of 5 or 9 beams represents a trade‐off between treatment time and plan quality. It has been shown that for more than 7–9 beams, the choice of beam directions does not matter for coplanar IMRT deliveries.(25,26)
Optimization was performed in “beamlet” mode with approximately 250 iterations performed before the final dose was calculated. After 250 iterations, the cost function has converged for all plans. The final dose matrix was calculated with the Analytical Anisotropic Algorithm (AAA, version 8.6.10) using a voxel size of
. The CPU within Eclipse is a dual Intel Xeon quad‐core processor running at 2.50 GHz.
D. RapidArc planning
RapidArc plans were created in the same Eclipse system, which supports planning with more than one arc. Single 360° rotation RapidArc (RA1) and dual 360° rotation RapidArc (RA2) treatment plans were created with the collimator rotated to 45°. The implementation of RapidArc with either 1 or 2 arcs represents a trade‐off between plan quality and treatment time whereby single arc plans are expected to be deliverable in a short period of time. The consequence of adding an additional arc may improve plan quality with an increase in the treatment time. A technique was developed to initialize the MLC positions of all arc segments to the PTV outline minus the OARs. In this version of the RapidArc planning system, the optimization appears to switch to the next MR level before the objective function has converged for a given level. The switch can be delayed by the user, and this was done progressively for 1 minute at MR1 up to 5 minutes at MR5. Upon completion of the “Arc Optimization”, the progress of the optimization was saved and the optimization was then “Continued” at resolution level five for 5 additional minutes. The final dose calculation was computed with AAA and a voxel size of
. The hardware specifications for the Eclipse system are the same as described in the previous section.
E. Tomotherapy planning
Tomotherapy treatment plans were created using the TomoTherapy planning station version 3.1.2.3. In tomotherapy planning, the field width is defined as the axial thickness of the fan beam, the pitch is defined as the couch travel distance for a complete gantry rotation relative to the axial beam width at the axis of rotation, and the modulation factor is defined as the maximum leaf opening time divided by the average leaf opening time. The pitch factors were chosen to reduce the thread effect on the final dose distribution and were chosen to be 0.86/n, where n is an integer.27
Treatment planning consisted of performing a full beamlet dose calculation following 250 optimization iterations. After 250 iterations the cost function has converged for all plans. The tomotherapy plans (TOMO) were created with a fan beam width of 2.5 cm at isocenter, a pitch of 0.287, and a modulation factor of 2.5. The dimensions of the dose voxels of tomotherapy treatment plans were
. The specifications for the tomotherapy clusters are 16 Intel dual processors running at 2.4 GHz.
F. Plan quality
In this study, the primary metric of plan quality was the ability to meet dose volume criteria during optimization as a binary metric as described below in F.1. The ability to meet DVH objectives can be used along with dose conformity and homogeneity and mean dose to the OARs as described in sections F.2 and F.3 below, respectively.
F.1 Dose volume evaluation
Upon completion of optimization, all plans were brought into the Computational Environment for Radiotherapy Research (CERR version 3.1) for dosimetric evaluation.28 All plans were optimized such that 95% of the volume of the PTV received a dose of 61 Gy.
The minimum dose is reported as the dose to 99.5% of the PTV and the maximum dose is reported as the dose to 0.5% of the PTV. Dose‐volume criteria, listed in Table 1, are reported as the percent volume receiving dose D,
and dose to a minimum percent volume V
. Figure 3 is a graphical representation of how the dose‐volume constraints are reported. A binary metric which includes the ability to meet DVH criteria was used for evaluation of treatment plans.

F.2 Conformity and homogeneity indices
The radiation conformity index (RCI) and dose homogeneity index (DHI) of the resultant treatment plans were computed using Eqs. 1 and 2, respectively. The RCI describes the volume within the PTV that receives the prescription dose or higher
divided by the volume within the external body contour that receives the prescription dose or higher
and provides a metric to determine how close the prescription isodose volume is spread within the PTV and normal tissue.
(1)The DHI describes the uniformity of the dose within the planning target volume and is a ratio of the minimum dose
to the maximum dose
.
(2)F.3 Mean dose to the OARs
The mean dose to each of the OARs was also recorded. The volume points that were desired were the dose to 50% of the volume and the maximum dose to the OAR; therefore, the mean dose should be correlated with the desired dose‐volume points.
G. Integral dose
The integral dose is reported as the sum of all dose voxels times their mass, as shown in Eq. 3. The phantom has unit density, so the equation can be simplified as shown below, where N is the number of voxels,
the mean dose to the body contour, and
the mass of a voxel. The simplest form of the equation is the mass of the body of the phantom
times the mean dose to the body of the phantom
.
(3) H. Volume receiving
and 
Some models of radiation carcinogenesis suggest that the dose‐response relationship is linear up until a dose of 6 Gy, where it then reaches a plateau.(29,30) The volumes receiving greater than 2 Gy and greater than 5 Gy would be important in this context and are reported for each treatment plan.
I. Estimation of treatment planning times
The treatment planning time is defined as the time from starting a plan until final optimization and dose calculation are completed. It is assumed that the plan can be delivered clinically and therefore would include selecting the relevant parameters, optimizing the delivery, and performing a final dose calculation.
J. Estimation of treatment delivery time
The estimated treatment delivery time will be defined as the time from first beam on until the last beam is turned off. For sliding window IMRT, it is the dose rate multiplied by the number of MU per field, plus the time that the gantry takes to rotate between successive fields, plus a parameter “delta” which takes into account the time for mode up, data transfer of the MLC delivery files, error in the estimated rotation time, and operator reaction time. The delta parameter was not added for the first beam because the data transfer and mode up happens before the first beam. Delta was determined (from 215 head and neck IMRT fields delivered at our center) as the difference between the theoretical beam‐on time at 400 MU/min plus gantry rotation at 360°/min and the actual delivery time as recorded by the record and verify system. For RapidArc, the estimated treatment delivery time is the sum of the time spent at each of the 177 segments within an arc. The time is the sum of all angular increment per segment divided by the gantry rotation rate. For multiple arc plans, the delta parameter was added for the second arc because all data transfer happens before the beam is turned on. The estimated treatment delivery times for tomotherapy are calculated by the planning system and recorded from the final plan report.
III. RESULTS
A. dose‐volume evaluation
The dose‐volume evaluation was performed, and data from all treatment plans is presented in Table 2. Representative DVH curves for the 9‐field IMRT, 2‐arc RapidArc, and tomotherapy plans are shown in Fig. 4. Figure 5 shows isodose lines taken for axial slices from the 9‐field IMRT, dual arc RapidArc, and tomotherapy dose distributions. Based on the data from Table 2, Phantom 1 had 0/4 optimization constraints met for IMRT5 and IMRT9, 1/4 constraints met for RA1 and RA2 and 2/4 for TOMO; Phantom 2 had 0/6 optimization constraints met for IMRT5, IMRT9 and RA1, 1/6 constraints met for RA2, and 4/6 for TOMO. For Phantom 3, all of the plans had 0/8 constraints met. Finally, for Phantom 4, only TOMO had 2/4 constraints met and the rest of the plans had 0/4 constraints met.
| PTV | OAR1 | OAR2 | OAR3 | |||||||||
| Min Dose | Max Dose | DVH ‐ Vol | DVH Dose | Max Dose | DVH ‐ Vol | DVH ‐ Dose | Max Dose | DVH ‐ Vol | DVH ‐ Dose | Max Dose | ||
| Phantom 1 | 60 | 65 | 50 | 13.3 | 20 | |||||||
| IMRT | 5 field | 59.2 | 68.9 | 82.1 | 16.0 | 29.3 | ||||||
| 9 field | 59.3 | 73.1 | 87.1 | 16.4 | 30.3 | |||||||
| RapidArc | 1 arc | 59.2 | 68.7 | 40.9 | 12.4 | 23.6 | ||||||
| 2 arc | 59.5 | 67.0 | 40.1 | 12.5 | 22.6 | |||||||
| Tomo | 59.1 | 67.2 | 15.1 | 9.4 | 17.0 | |||||||
| Phantom 2 | 60.0 | 65.0 | 50.0 | 10.0 | 20.0 | 50.0 | 10.0 | 20.0 | ||||
| IMRT | 5 field | 58.2 | 69.3 | 97.7 | 13.2 | 28.6 | 63.8 | 25.3 | 27.6 | |||
| 9 field | 57.0 | 69.2 | 90.4 | 13.1 | 31.6 | 64.8 | 25.7 | 29.5 | ||||
| RapidArc | 1 arc | 58.1 | 68.7 | 58.5 | 10.3 | 23.4 | 52.3 | 10.2 | 24.1 | |||
| 2 arc | 58.3 | 68.0 | 50.2 | 10.0 | 23.1 | 47.3 | 9.8 | 22.9 | ||||
| Tomo | 59.9 | 66.1 | 42.6 | 9.5 | 17.4 | 33.9 | 8.8 | 17.9 | ||||
| Phantom 3 | 60.0 | 65.0 | 50.0 | 10.0 | 20.0 | 50.0 | 10.0 | 20.0 | 50.0 | 15.0 | 30.0 | |
| IMRT | 5 field | 58.8 | 70.4 | 79.6 | 12.9 | 28.4 | 66.1 | 13.3 | 35.4 | 66.9 | 19.5 | 44.7 |
| 9 field | 58.3 | 70.1 | 84.8 | 14.3 | 31.3 | 69.4 | 13.6 | 40.1 | 69.7 | 18.7 | 46.8 | |
| RapidArc | 1 arc | 58.3 | 69.0 | 94.9 | 13.4 | 24.7 | 81.9 | 13.1 | 28.6 | 91.3 | 21.7 | 44.2 |
| 2 arc | 58.3 | 69.0 | 90.7 | 13.5 | 26.1 | 88.1 | 13.2 | 27.8 | 93.8 | 21.2 | 43.6 | |
| Tomo | 56.7 | 66.7 | 99.9 | 15.1 | 24.6 | 96.8 | 13.5 | 25.2 | 83.0 | 21.5 | 43.6 | |
| Phantom 4 | 60.0 | 65.0 | 50.0 | 15.0 | 30.0 | |||||||
| IMRT | 5 field | 55.0 | 75.3 | 86.5 | 23.3 | 41.4 | ||||||
| 9 field | 57.6 | 69.5 | 87.6 | 21.6 | 39.4 | |||||||
| 2 arc | 59.0 | 66.5 | 68.3 | 16.9 | 33.8 | |||||||
| Tomo | 57.4 | 65.6 | 28.7 | 11.2 | 29.5 | |||||||


B. Radiation conformity index, dose homogeneity index
A complete list of the radiation conformity and dose homogeneity indices for all plans is included in Table 3. The average RCI values were
for IMRT,
for RapidArc, and
for tomotherapy. The average DHI values were
for IMRT,
for RapidArc, and
for tomotherapy.
| RCI | DHI | Mean OAR1 | Mean OAR2 | Mean OAR3 | Integral Dose | ![]() |
![]() |
|||
|---|---|---|---|---|---|---|---|---|---|---|
| Phantom | IMRT | 5 field | 0.855 | 0.860 | 17.32 | 89.0 | 4148 | 3275 | ||
| 1 | 9 field | 0.913 | 0.811 | 17.78 | 87.8 | 4151 | 3425 | |||
| RapidArc | 1 arc | 0.890 | 0.862 | 13.65 | 85.8 | 4338 | 3379 | |||
| 2 arc | 0.914 | 0.888 | 13.53 | 85.7 | 4311 | 3381 | ||||
| Tomo | 0.706 | 0.880 | 10.47 | 106.9 | 4884 | 4281 | ||||
| Phantom | IMRT | 5 field | 0.876 | 0.839 | 14.66 | 13.88 | 59.2 | 2617 | 2051 | |
| 2 | 9 field | 0.921 | 0.824 | 14.90 | 14.28 | 59.0 | 2662 | 2101 | ||
| RapidArc | 1 arc | 0.911 | 0.846 | 11.63 | 11.68 | 58.0 | 3019 | 2181 | ||
| 2 arc | 0.926 | 0.857 | 11.20 | 11.19 | 57.2 | 2970 | 2158 | |||
| Tomo | 0.793 | 0.906 | 10.07 | 9.93 | 75.2 | 3507 | 2995 | |||
| Phantom | IMRT | 5 field | 0.758 | 0.835 | 14.55 | 15.16 | 21.26 | 59.1 | 2630 | 2035 |
| 3 | 9 field | 0.783 | 0.832 | 15.86 | 16.17 | 21.46 | 58.3 | 2633 | 2031 | |
| RapidArc | 1 arc | 0.904 | 0.844 | 14.44 | 14.56 | 24.27 | 58.6 | 3104 | 2226 | |
| 2 arc | 0.911 | 0.846 | 14.43 | 14.71 | 23.81 | 58.2 | 3060 | 2223 | ||
| Tomo | 0.847 | 0.850 | 15.74 | 14.77 | 23.23 | 73.5 | 3510 | 2977 | ||
| Phantom | IMRT | 5 field | 0.769 | 0.730 | 24.34 | 148.3 | 4991 | 3477 | ||
| 4 | 9 field | 0.912 | 0.829 | 23.06 | 142.3 | 4833 | 3450 | |||
| RapidArc | 1 arc | 0.904 | 0.826 | 20.24 | 142.1 | 5562 | 3915 | |||
| 2 arc | 0.940 | 0.887 | 18.81 | 141.2 | 5629 | 3915 | ||||
| Tomo | 0.900 | 0.874 | 13.26 | 158.9 | 5367 | 4195 |
C. Mean dose to the OARs
The mean dose values for each phantom and delivery technique are listed in Table 3. The general trend regarding the mean dose to the OARs for Phantoms 1, 2 and 4 is
.
D. Integral dose and volumes receiving in excess of 2 and 5 Gy
The integral doses, in Joules, are listed in Table 3. Tomotherapy delivers on average 20% more integral dose than IMRT and RapidArc. The volumes receiving in excess of 2 and 5 Gy are listed in Table 3. The general trend is that tomotherapy has the highest volume receiving in excess of 2 and 5 Gy, followed by RapidArc, with IMRT plans having the lowest.
E. Treatment plan time and estimated treatment delivery time
A summary of the planning time and estimated treatment delivery time is shown in Table 4. The planning time is on average 7.5 minutes for IMRT, 48 minutes for RapidArc, and 59 minutes for tomotherapy. The estimated delivery time is on average 4.8 minutes for IMRT, 2.2 minutes for RapidArc, and 3.5 minutes for tomotherapy. The average delta value per IMRT field was determined to be
seconds per beam based on 215 head and neck IMRT fields.
| Phantom 1 | Phantom 2 | Phantom 3 | Phantom 4 | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Plan | Deliver | Plan | Deliver | Plan | Deliver | Plan | Deliver | ||
| IMRT | 5 field | 6.8 | 3.8 | 6.7 | 3.9 | 6.7 | 4.0 | 7.3 | 4.3 |
| 9 field | 8.1 | 5.2 | 7.5 | 5.3 | 7.8 | 5.8 | 9.1 | 5.9 | |
| RapidArc | 1 arc | 45.0 | 1.4 | 38.0 | 1.4 | 30.0 | 1.7 | 53.0 | 1.7 |
| 2 arc | 60.0 | 2.8 | 47.0 | 2.8 | 50.0 | 2.8 | 62.0 | 2.8 | |
| Tomo | 45.0 | 4.3 | 44.0 | 2.8 | 44.0 | 2.8 | 101.0 | 3.9 | |
IV. DISCUSSION
For all phantoms, tomotherapy was able to meet the most optimization criteria (2/4 for P1, 3/6 for P2, 2/4 for P4), followed by RapidArc (1/4 for P1, 1/6 for P2 and 0/4 for P3), followed by IMRT (0/4 for P1, 0/6 for P2 and 0/4 for P4). These results are generalized for treatment techniques and are mostly independent of the number of beams/arcs used. Tomotherapy plans were able to produce the most homogeneous dose to the PTV as an average for all phantoms
. This improvement in ability to meet DVH criteria comes with some costs including: longer planning time
, longer estimated treatment times
, lower conformity index
, and higher integral dose (approximately 1.2 times values attained for IMRT or RapidArc).
For Phantom 3, none of the treatment plans was able to meet any of the dose volume criteria. However, tomotherapy was able to provide the most dose homogeneity for that plan
, but less conformity
. There was no trend regarding the mean dose to the OARS for Phantom 3.
This study demonstrates that treatment techniques differ in terms of the trade‐offs between treatment planning time, treatment delivery time, and overall plan quality. It is clear that sliding window IMRT treatment plans can be created in a much shorter period of time as compared to either RapidArc or tomotherapy, and that RapidArc (either single or dual beam) has the lowest estimated treatment delivery time compared to both IMRT and tomotherapy. With respect to plan quality, it appears as though tomotherapy can meet the most dose‐volume criteria and can, on average, produce plans with the most homogeneity within the target volume. The average conformity indices show that tomotherapy is worse than both IMRT and RapidArc, even though tomotherapy is thought to have much more flexibility in shaping dose distribution during optimization.22 Our result can be explained by at least two reasons. First, the two treatment planning systems (TomoTherapy and Eclipse) may have different criteria for considering whether voxels are either inside or outside of specific contours. Second, in order to deliver full dose within the target with tomotherapy, the “ramp‐up” region needs to be at least as thick as the fan beam thickness, leading to additional dose superior and inferior to the target volume.31 Superior‐inferior dose profiles through the geometric center of Phantom 4 are shown in Fig. 6 for 9‐field IMRT, 2‐arc RapidArc, and tomotherapy to illustrate the ramp‐up effect.

to
.
In this study, a single pass approach to inverse planning was undertaken as opposed to an approach where the optimization parameters are changed over multiple iterations in order to produce the plan that meets the most DVH criteria. A more robust methodology would be required to definitively rank the plan quality for IMRT, RA and Tomo. For example, by iteratively adjusting the optimization parameters the expert planner can obtain better plans, as compared to a single pass approach. In addition, it would be useful to recruit expert planners for each treatment modality. It remains to be seen what the rank order of treatment modalities would be from such a study because IMRT plans can be created in a very short period of time and many more optimization parameter changes can be exercised for IMRT as compared to RA or Tomo for a fixed planning time.
An additional limitation of a treatment planning study is that the intercomparison of data from different treatment planning systems may not be fair due to a number of complicating issues. One issue is termed the “weight paradox” by Deasy, whereby the optimal choices for the relative weights of different PTV and OAR optimization criteria are not known and may take many iterations of trial and error to determine.32 Choosing DVH criteria that are too constraining on the OAR compromises the PTV coverage, or vice versa, and may end up producing a plan that the user does not desire.32 These dosimetric obstacles have been addressed by two different approaches including multi‐criteria optimization(33–35) and Pareto optimization,(36,37) and when implemented in commercial treatment planning systems may be able to provide improved results. Another challenging issue is that treatment planning systems from individual manufacturers may define the objective function and weighting values for the PTV and OAR differently. This information is generally not disclosed to the user and is difficult to interpret.
One limitation in this study is that the phantoms were essentially 2D axial phantoms extended in the third dimension and do not represent contours that would be seen in a clinical environment. It is uncertain whether the TomoTherapy system would have significant difficulty in planning on phantoms with PTV and OAR contours that change shapes in the superior‐inferior direction. As mentioned in the paper by Bortfeld and Webb, within tomotherapy there is an efficiency trade‐off when planning in the superior‐inferior direction, this trade‐off involves the choices of fan beam width, modulation factor, and pitch.23 An additional study is necessary to understand how these trade‐offs impact plan quality and treatment efficiency for structure sets that vary in shape in the superior‐inferior direction.
Finally, it is difficult to know that the optimization algorithm has converged to a local minimum or how close the solution is to the global minimum. Although there are visual displays which indicate to the user how many iterations have elapsed, there isn't a criteria to stop the optimization if the optimization has converged to a set criteria such as 1% difference in objective function value change over 100 iterations. If such stopping criteria were available, then statements could be made regarding the convergence of the respective optimization algorithms.
Additional work is necessary to further understand the relative merits of differing delivery technologies for clinical geometries.
V. CONCLUSIONS
Advanced radiation therapy delivery techniques each have their own relative merits. For the four phantoms investigated in this study, 5‐ and 9‐field IMRT treatments can be planned in the shortest time and can be delivered with the lowest integral dose. Single and dual arc RapidArc plans can be delivered in the shortest period of time and can provide the most conformal deliveries to the PTV. Finally, tomotherapy is capable of meeting most of the planning objectives and can provide the most uniform dose to the PTV.
ACKNOWLEDGEMENTS
We would like to thank the London Regional Cancer Program (Drs. Jeff Chen and Slav Yartsev) for allowing us to use the TomoTherapy planning station for this study.
Number of times cited: 47
- Shih-Ming Hsu, Yuan-Chun Lai, Chien-Chung Jeng and Chia-Ying Tseng, Dosimetric comparison of different treatment modalities for stereotactic radiotherapy, Radiation Oncology, 12, 1, (2017).
- Kazuki Kubo, Hajime Monzen, Kentaro Ishii, Mikoto Tamura, Ryu Kawamorita, Iori Sumida, Hirokazu Mizuno and Yasumasa Nishimura, Dosimetric comparison of RapidPlan and manually optimized plans in volumetric modulated arc therapy for prostate cancer, Physica Medica, 44, (199), (2017).
- Daniel Buergy, Abdul Wahab M. Sharfo, Ben J. M. Heijmen, Peter W. J. Voet, Sebastiaan Breedveld, Frederik Wenz, Frank Lohr and Florian Stieler, Fully automated treatment planning of spinal metastases – A comparison to manual planning of Volumetric Modulated Arc Therapy for conventionally fractionated irradiation, Radiation Oncology, 12, 1, (2017).
- Shabbir Ahamed, Navin Singh, Deleep Gudipudi, Suneetha Mulinti, Anil Talluri, Bhudevi Soubhagya and Madhusudhana Sresty, Assessment of monitor unit limiting strategy using volumetric modulated arc therapy for cancer of hypopharynx, Physica Medica, 35, (73), (2017).
- Ahmad Amoush, Huang Long, Laxmi Subedi, Peng Qi, Toufik Djemil and Ping Xia, Dosimetric effect of multileaf collimator leaf width on volumetric modulated arc stereotactic radiotherapy for spine tumors, Medical Dosimetry, 42, 2, (111), (2017).
- Lionel Perrier, Magali Morelle, Pascal Pommier, Pierre Boisselier, Bernard Coche-Dequeant, Olivier Gallocher, Marc Alfonsi, Etienne Bardet, Michel Rives, Valentin Calugaru, Enrique Chajon, Georges Noel, Hinda Mecellem, David Pérol, Sophie Dussart and Philippe Giraud, Cost Analysis of Complex Radiation Therapy for Patients With Head and Neck Cancer, International Journal of Radiation Oncology*Biology*Physics, 10.1016/j.ijrobp.2016.02.013, 95, 2, (654-662), (2016).
- Philip McGeachy, Jose Eduardo Villarreal-Barajas, Yuriy Zinchenko and Rao Khan, Modulated photon radiotherapy (XMRT): an algorithm for the simultaneous optimization of photon beamlet energy and intensity in external beam radiotherapy (EBRT) planning, Physics in Medicine and Biology, 61, 4, (1476), (2016).
- Zhen Tian, Fei Peng, Michael Folkerts, Jun Tan, Xun Jia and Steve B. Jiang, Multi‐GPU implementation of a VMAT treatment plan optimization algorithm, Medical Physics, 42, 6Part1, (2841-2852), (2015).
- KR Muralidhar, Bhudevi Soubhagya and Shabbir Ahmed, Intensity modulated radiotherapy versus volumetric modulated arc therapy in breast cancer: A comparative dosimetric analysis, International Journal of Cancer Therapy and Oncology, 3, 2, (3210), (2015).
- P. Fenoglietto, C. Bourgier, O. Riou, C. Lemanski and D. Azria, Impact de la modulation d’intensité dans l’irradiation des aires ganglionnaires du cancer du sein, Cancer/Radiothérapie, 19, 4, (265), (2015).
- Kristoffer Petersson, Per Nilsson, Per Engström, Tommy Knöös and Crister Ceberg, Multi-modality optimisation in radiotherapy treatment planning using composite objective values, Acta Oncologica, 54, 4, (557), (2015).
- Esther Wild, Mark Bangert, Simeon Nill and Uwe Oelfke, Noncoplanar VMAT for nasopharyngeal tumors: Plan quality versus treatment time, Medical Physics, 42, 5, (2157-2168), (2016).
- Sonja Katayama, Matthias F Haefner, Angela Mohr, Kai Schubert, Dieter Oetzel, Juergen Debus and Florian Sterzing, Accelerated tomotherapy delivery with TomoEdge technique, Journal of Applied Clinical Medical Physics, 16, 2, (33-42), (2015).
- Prabhjot Juneja, Andrew Kneebone, Jeremy T. Booth, David I. Thwaites, Ramandeep Kaur, Emma Colvill, Jin A. Ng, Paul J. Keall and Thomas Eade, Prostate motion during radiotherapy of prostate cancer patients with and without application of a hydrogel spacer: a comparative study, Radiation Oncology, 10, 1, (2015).
- Lotte S. Fog, Deborah Schut, Per Sjøgren and Marianne C. Aznar, The potential role of modern radiotherapy techniques in the treatment of malignant spinal cord compression: a dose planning study, Journal of Radiotherapy in Practice, 14, 04, (418), (2015).
- Sarah E. Goers and Caroline M. Fadipe, Novel applications of TomoTherapy in the treatment of metastases, Radiography, 21, 2, (203), (2015).
- Ahmad Amoush, Allison Dalton, Bryan Rabatic, Ke Huang and Ahmad Al-Basheer, Volumetric modulated arc therapy for spine SBRT patients to reduce treatment time and intrafractional motion, International Journal of Cancer Therapy and Oncology, 3, 2, (03026), (2015).
- Geert De Kerf, Dirk Van Gestel, Lobke Mommaerts, Danielle Van den Weyngaert and Dirk Verellen, Evaluation of the optimal combinations of modulation factor and pitch for Helical TomoTherapy plans made with TomoEdge using Pareto optimal fronts, Radiation Oncology, 10, 1, (2015).
- S Chakraborty, V M Patil, S Babu, G Muttath and S K Thiagarajan, Locoregional recurrences after post-operative volumetric modulated arc radiotherapy (VMAT) in oral cavity cancers in a resource constrained setting: experience and lessons learned, The British Journal of Radiology, 88, 1048, (20140795), (2015).
- Y. M. Yang, M. Geurts, J. B. Smilowitz, E. Sterpin and B. P. Bednarz, Monte Carlo simulations of patient dose perturbations in rotational‐type radiotherapy due to a transverse magnetic field: A tomotherapy investigation, Medical Physics, 42, 2, (715-725), (2015).
- A.T. Falk, P. Fenoglietto, D. Azria and C. Bourgier, Nouvelles techniques de radiothérapie externe : exemple de prise en charge dans les cancers du sein, Cancer/Radiothérapie, 18, 5-6, (480), (2014).
- Elizabeth Forde, Regina Bromley, Andrew Kneebone and Thomas Eade, A class solution for volumetric-modulated arc therapy planning in postprostatectomy radiotherapy, Medical Dosimetry, 39, 3, (261), (2014).
- Mirza Ali, Muntimadugu Babaiah, Nagaraju Madhusudhan and Geomcy George, Comparative dosimetric analysis of IMRT and VMAT (RapidArc) in brain, head and neck, breast and prostate malignancies, International Journal of Cancer Therapy and Oncology, 3, 1, (03019), (2014).
- R.M. Howell and S.F. Kry, Modulated Arc Therapy Planning, Comprehensive Biomedical Physics, 10.1016/B978-0-444-53632-7.00915-1, (395-400), (2014).
- Leonie Rudofsky, Eleanor Aynsley, Sebastian Beck, Kai Schubert, Gregor Habl, Sonja Krause, Jürgen Debus and Florian Sterzing, Lung and liver SBRT using helical tomotherapy — a dosimetric comparison of fixed jaw and dynamic jaw delivery, Journal of Applied Clinical Medical Physics, 15, 3, (114-121), (2014).
- Hany Ammar, Eman Eldebawy, Ehab Maarouf, Wafaa Khalil and MS Zaghloul, Evaluation of the peripheral dose and the conformity index for three stereotactic radiotherapy techniques: Arcs, noncoplanar fixed fields and intensity modulation, International Journal of Cancer Therapy and Oncology, 2, 4, (02042), (2014).
- Craig A. Elith, Shane E. Dempsey and Helen M. Warren‐Forward, Comparing four volumetric modulated arc therapy beam arrangements for the treatment of early‐stage prostate cancer, Journal of Medical Radiation Sciences, 61, 2, (91-101), (2014).
- David Craft, Dávid Papp and Jan Unkelbach, Plan averaging for multicriteria navigation of sliding window IMRT and VMAT, Medical Physics, 41, 2, (2014).
- G. Heilemann, B. Poppe and W. Laub, On the sensitivity of common gamma‐index evaluation methods to MLC misalignments in Rapidarc quality assurance, Medical Physics, 40, 3, (2013).
- Hani Ashamalla, Ajay Tejwani, Ioannis Parameritis, Uma Swamy, Pei Ching Luo, Adel Guirguis and Amir Lavaf, Comparison study of intensity modulated arc therapy using single or multiple arcs to intensity modulated radiation therapy for high-risk prostate cancer, Radiation Oncology Journal, 31, 2, (104), (2013).
- Ulrich W. Langner, Janelle A. Molloy, John F. Gleason and Jonathan M. Feddock, A feasibility study using TomoDirect for craniospinal irradiation, Journal of Applied Clinical Medical Physics, 14, 5, (104-114), (2013).
- Monica W.K. Kan, Lucullus H.T. Leung and Peter K.N. Yu, The performance of the progressive resolution optimizer (PRO) for RapidArc planning in targets with low‐density media, Journal of Applied Clinical Medical Physics, 14, 6, (205-221), (2013).
- Mu-Han Lin, Robert A Price, Jinsheng Li, Shengwei Kang, Jie Li and C-M Ma, Investigation of pulsed IMRT and VMAT for re-irradiation treatments: dosimetric and delivery feasibilities, Physics in Medicine and Biology, 58, 22, (8179), (2013).
- Neelam Tyagi, Kai Yang, Raminder Sandhu, Di Yan, Sean S. Park, Peter Y. Chen and Brian Marples, External beam pulsed low dose radiotherapy using volumetric modulated arc therapy: Planning and delivery, Medical Physics, 40, 1, (2012).
- J A Ng, J Booth, P Poulsen, Z Kuncic and P J Keall, Estimation of effective imaging dose for kilovoltage intratreatment monitoring of the prostate position during cancer radiotherapy, Physics in Medicine and Biology, 58, 17, (5983), (2013).
- James K Crocker, Jin Aun Ng, Paul J Keall and Jeremy T Booth, Measurement of patient imaging dose for real-time kilovoltage x-ray intrafraction tumour position monitoring in prostate patients, Physics in Medicine and Biology, 57, 10, (2969), (2012).
- David Craft, Dualta McQuaid, Jeremiah Wala, Wei Chen, Ehsan Salari and Thomas Bortfeld, Multicriteria VMAT optimization, Medical Physics, 39, 2, (686-696), (2012).
- M Ahamed Badusha and CK McGarry, Practical collimator optimization in the management of prostate IMRT planning: A feasibility study, Journal of Radiotherapy in Practice, 11, 02, (107), (2012).
- Gregory T. Betzel, Byong Yong Yi, Ying Niu and Cedric X. Yu, Is RapidArc more susceptible to delivery uncertainties than dynamic IMRT?, Medical Physics, 39, 10, (5882-5890), (2012).
- Rasmus Bokrantz, Multicriteria optimization for volumetric‐modulated arc therapy by decomposition into a fluence‐based relaxation and a segment weight‐based restriction, Medical Physics, 39, 11, (6712-6725), (2012).
- Monica W.K. Kan, Wicger Wong, Lucullus H.T. Leung, Peter K.N. Yu, Ronald W.K. So and Ashley C.K. Cheng, A comprehensive dosimetric evaluation of using RapidArc volumetric‐modulated arc therapy for the treatment of early‐stage nasopharyngeal carcinoma, Journal of Applied Clinical Medical Physics, 13, 6, (189-202), (2012).
- Chris Beltran, Jonathan Gray and Thomas E. Merchant, Intensity-Modulated Arc Therapy for Pediatric Posterior Fossa Tumors, International Journal of Radiation Oncology*Biology*Physics, 82, 2, (e299), (2012).
- Kai Yang, Di Yan and Neelam Tyagi, Sensitivity analysis of physics and planning SmartArc parameters for single and partial arc VMAT planning, Journal of Applied Clinical Medical Physics, 13, 6, (34-45), (2012).
- Craig Elith, Shane E. Dempsey, Naomi Findlay and Helen M. Warren-Forward, An Introduction to the Intensity-modulated Radiation Therapy (IMRT) Techniques, Tomotherapy, and VMAT, Journal of Medical Imaging and Radiation Sciences, 42, 1, (37), (2011).
- David Jolly, Dineli Alahakone and Juergen Meyer, A RapidArc planning strategy for prostate with simultaneous integrated boost, Journal of Applied Clinical Medical Physics, 12, 1, (35-49), (2010).
- Mike Oliver, Karl Bush, Sergei Zavgorodni, Will Ansbacher and Wayne A. Beckham, Understanding the impact of RapidArc therapy delivery errors for prostate cancer, Journal of Applied Clinical Medical Physics, 12, 3, (32-43), (2011).
- Spencer Martin, Jeff Z. Chen, A. Rashid Dar and Slav Yartsev, Dosimetric comparison of helical tomotherapy, RapidArc, and a novel IMRT & Arc technique for esophageal carcinoma, Radiotherapy and Oncology, 101, 3, (431), (2011).






